No single ISO standard covers medical devices as a whole. Instead, FDA submissions and inspections draw on a set of standards that each cover one part of the product or quality system: ISO 13485 for the quality management system, ISO 14971 for risk management, the ISO 10993 series for biocompatibility, IEC 62304 for software, and several more, each with its own role.
Conformance to most consensus standards is voluntary. The main exception is ISO 13485:2016, which the Quality Management System Regulation (21 CFR Part 820) incorporates by reference, so its requirements are part of FDA regulation. For other standards, FDA maintains a recognized consensus standards database: a list of standards, by specific edition, against which FDA will accept a manufacturer's Declaration of Conformity in a premarket submission. Citing a recognized standard correctly can reduce the supporting documentation needed for that part of a submission.
Standards Most Relevant to an FDA Submission or Inspection
| Standard | What it covers | Where it matters |
|---|---|---|
| ISO 13485 | Quality management systems for medical device organizations | FDA's QMSR (21 CFR Part 820) incorporates ISO 13485:2016 by reference (implementation guide →) |
| ISO 14971 | Risk management applied throughout the device lifecycle | Referenced throughout design controls, submissions, and inspections (what FDA expects →) |
| ISO 10993 series | Biological evaluation of materials that contact the body | Biocompatibility section of a 510(k), PMA, or De Novo submission (what it is →) |
| IEC 62304 | Software lifecycle processes for medical device software | Software documentation for any device with embedded or standalone software (practical guide →) |
| IEC 62366-1 | Usability engineering applied to medical devices | Human factors section of a submission for devices with a meaningful use-error risk (usability engineering →) |
| IEC 60601 series | Basic safety and essential performance of medical electrical equipment | Electromechanical device submissions |
| ISO 11135 | Validation of ethylene oxide sterilization | Sterilization validation for EO-sterilized devices (sterilization validation guide →) |
| ISO 11137 | Validation of radiation sterilization | Sterilization validation for radiation-sterilized devices (sterilization validation guide →) |
| ISO 11607 | Packaging for terminally sterilized medical devices | Sterile barrier system validation ahead of a 510(k) or PMA (packaging design & validation →) |
| ISO 15223-1 | Symbols for use in medical device labeling | Labeling review |
| ISO 14155 | Good clinical practice for clinical investigation of medical devices | IDE and clinical study design |
ISO 13485 Key Requirements
ISO 13485:2016 sets its requirements in five clauses, 4 through 8:
- Quality management system (clause 4). A documented quality system, including a quality manual, a medical device file for each device type or family, and control of documents and records.
- Management responsibility (clause 5). A documented quality policy, defined responsibilities and authorities, and regular management review of the quality system's performance.
- Resource management (clause 6). Competent, trained personnel and a controlled work environment appropriate to the product.
- Product realization (clause 7). Design and development controls, purchasing controls over suppliers, production and process controls, and identification and traceability through the product lifecycle.
- Measurement, analysis and improvement (clause 8). Internal audits, monitoring of customer feedback and complaints, control of nonconforming product, and corrective and preventive action (CAPA).
The standard applies a risk based approach throughout and requires risk management during product realization. ISO 14971 is the standard normally used to meet that requirement. See the full ISO 13485 implementation guide for how these requirements translate into an actual quality manual and procedure set.
How FDA Recognizes Consensus Standards
FDA maintains a recognized consensus standards database for medical devices: a searchable list, by standards organization, standard number, and FDA specialty task group area (including a dedicated Biocompatibility area), of the standards FDA will accept a Declaration of Conformity against in a premarket submission. Recognition is standard-by-standard and edition-specific: FDA recognizes a particular edition of a standard, not the standard in the abstract, and the recognized edition can change over time as standards are revised. A manufacturer citing a standard in a submission should confirm the currently recognized edition directly in FDA's database rather than assume it from the standard's own publication date.
Building or Updating a Quality System Against These Standards?
ADB's regulatory team reviews a quality management system against ISO 13485:2016 and FDA QMSR, starting with a Stage 1 documentation and forms gap assessment.
See QMS Gap Assessment Book a Free ConsultFrequently Asked Questions
What are the ISO standards for medical devices?
There is no single ISO standard that covers all medical devices. Instead, a set of standards each cover a different part of the product and quality system, the most common being ISO 13485 for quality management, ISO 14971 for risk management, the ISO 10993 series for biocompatibility, IEC 62304 for software, IEC 62366-1 for usability, and ISO 11135/11137/11607 for sterilization and packaging.
What are the key requirements of ISO 13485?
ISO 13485:2016 requires a documented quality management system covering quality system documentation, management responsibility, resource management, product realization (including design controls and purchasing controls), and measurement, analysis, and improvement (including CAPA and complaint handling), with risk management applied throughout the product lifecycle.
How does FDA recognize consensus standards like these?
FDA maintains a recognized consensus standards database that lists the specific standards, by edition, to which FDA will accept a manufacturer's Declaration of Conformity in a premarket submission. Recognition is standard-by-standard and edition-specific, so a manufacturer should confirm the current recognized edition in FDA's database rather than assume it from the standard's publication date.
Is ISO 13485 certification the same as FDA QMSR compliance?
No, though the two are closely aligned. ISO 13485:2016 is a voluntary, third-party-certifiable standard; the FDA Quality Management System Regulation (QMSR, 21 CFR Part 820) incorporates ISO 13485:2016 by reference but is a binding US regulation enforced through FDA inspection, with some additional FDA-specific record requirements layered on top.